Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that would be felt by the patient as uncomfortable or painful. SJMs routine testing, monitoring, and analysis of manufacturing processes identified three lots of Eon Mini IPGs that the internal battery has the potential to come into contact with the internal microcontroller board.
Source facts
Action
ST. Jude Medical sent an important Medical Device Advisory letter dated September 4, 2012 to all affected customers. The letter identified the affected product, problem and recommended actions to be taken. For questions contact your St. Jude Medical Neuromodulation Division Representative or call the Technical Services Department at 1-800-727-7846.
Initiated
2012-09-04
Root cause
Nonconforming Material/Component
Terminated
2014-05-12
Distribution
US Nationwide Distribution
Recall number
Z-0025-2013
Recall status
Terminated
CDRH record ID
112696
FDA classified
2012-10-09
Recall event number
63154