Z-0025-2024: REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64

Recall record Z-0025-2024 for Becton Dickinson Infusion Therapy Systems Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0025-2024
Company in FDA record
Becton Dickinson Infusion Therapy Systems Inc.
FDA classification date
Oct 3, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.

Source facts

Action
On August 29, 2023, Becton Dickinson issued an "Urgent Medical Device Recall" notification via. FedEx. Becton Dickinson asked consignees to take the following actions: 1. Immediately review your inventory for Catalog Number 386862, Lot number 2238464. Destroy all unused product subject to the recall following your institution's process for destruction. 2. For indwelling catheters, no further action is needed. 3. If a defective product was used on a patient and the used needle was disposed of without the clinician suffering a needlestick injury, no further action is necessary to further protect the patient or the clinician. 4. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch, Phone: 1-800-FDA 1088 (1-800-332-1088), Mail: MedWatch, HF-2, FDA, 5600 Fisher.s Lane, Rockville, MD 20852-978
Initiated
2023-08-29
Root cause
Process change control
Distribution
US Nationwide distribution in the states of CA, FL, GA, MA, MN, MO, MS, NJ, OH, TX.
Recall number
Z-0025-2024
Recall status
Open, Classified
CDRH record ID
203044
FDA classified
2023-10-03
Recall event number
92970

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗