Open original FDA record ↗Recall record
- FDA identifier
- Z-0026-2007
- Company in FDA record
- Extended Care Air Therapy Systems Inc
- FDA classification date
- Oct 14, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Component defect- Following assembly and distribution, two (2) bolts may bend in the bed frame, which may allow the bed to tip into the wall
Source facts
- Action
- The firm sent a Product Advisory Letter to their distributor/customers on 4/5/2006. The letter instructs the distributors as to the correct way to handle the beds during movement, loading, and unloading onto delivery trucks. The letter also urges the distributors to inspect the bed frame units in inventory at their locations for the presence of any bending, or damage to the four frame bolts. The letter states that the company will replace the bolts as soon as possible when ever necessary.
- Initiated
- 2006-04-05
- Root cause
- Other
- Terminated
- 2012-09-21
- Distribution
- The recalled medical beds were distributed to six (6) wholesale/distributor customers located in the following states: KS, TN, MI, PA, and LA.
- Recall number
- Z-0026-2007
- Recall status
- Terminated
- CDRH record ID
- 48047
- FDA classified
- 2006-10-14
- Recall event number
- 36125