Open original FDA record ↗Recall record
- FDA identifier
- Z-0026-2010
- Company in FDA record
- Hospira Inc
- FDA classification date
- Oct 6, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Source facts
- Action
- Firm initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and an Urgent: Device Field Correction letter to consignees was sent via UPS. The letter recommended that customers check their infusion pumps and spare parts in their inventory to determine if the specified AC Power Cords are in use at their facility; and, to inspect the affected cords for bent or cracked prongs and bridge connection, burnt plastic, or excessive wear and tear. Also, it was recommended that customers immediately discontinue use of the affected cords exhibiting the described characteristics and contact their local Hospira representative for assistance. The letter stated that customers may continue to use and monitor cords that are not exhibiting the described characteristics until replacement cords are sent to their facility. Hospira will contact customers to schedule replacement. Customers are to complete and return the attached Reply Form. Question should be directed to Hospira Customer Service at 1-800-241-4002.
- Initiated
- 2009-08-11
- Root cause
- Component design/selection
- Terminated
- 2010-11-30
- Distribution
- Worldwide Distribution
- Recall number
- Z-0026-2010
- Recall status
- Terminated
- CDRH record ID
- 84814
- FDA classified
- 2009-10-06
- Recall event number
- 52866