Skip to content

Public FDA record

Z-0026-2013: OrthoHelix Part #HCS-056-32 3.2MM Cannulated Hex Driver NON STERILE MAT: 465 SS The drivers are used

Recall record Z-0026-2013 for Orthohelix Surgical Designs Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0026-2013
Company in FDA record
Orthohelix Surgical Designs Inc
FDA classification date
Oct 9, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

The tips of the drivers are breaking while being used to drive the screws into patients.

Source facts

Action
The firm, OrthoHelix Surgical Designs, Inc., sent an "IMPORTANT RECALL NOTICE: Mini MaxTorque HCS-056-32 & HCS-056-38 Drivers" dated January 29, 2010 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to discontinue the use of these drivers immediately and return them to OrthoHelix Surgical Designs Inc, 1065 Medina Rd., Suite 500, Medina, OH 44256; ATTN: Driver Recall. Should you have any questions about system, call 330-247-1145 or the Director of Regulatory Affairs & Quality Assurance at 330-247-1444.
Initiated
2010-01-29
Root cause
Software design (manufacturing process)
Terminated
2013-02-20
Distribution
Nationwide distribution: USA including the states of CA, CO, FL, GA, IL, MO, NC, NH, NJ, NM, NY, OH, OR, PA, SD, TN, TX, VA, WA, and WI.
Recall number
Z-0026-2013
Recall status
Terminated
CDRH record ID
89070
FDA classified
2012-10-09
Recall event number
54725

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗