- Recall classification
- Class II
Reason stated in the source
The recalling firm has received reports of increased or decreased flows that have occurred in certain pumps.
Source facts
- Action
- BD sent an Urgent: Medical Device Recall Notification dated September 1, 2017. The customer letter will instruct the customers to remove the pump from service if it shows signs of infusion at an unexpected rate and not to use the affected devices in high risk areas if possible. The customer notification letter will be addressed to the Directors of Nursing, Risk Management, and Biomedical Engineering. Customers will be required to confirm receipt of the notification by returning the Recall Response Card to BD by postage-paid, self-addressed mail, fax, or email. For further questions, please call (888) 562-6018.
- Initiated
- 2017-08-09
- Root cause
- Unknown/Undetermined by firm
- Distribution
- Worldwide Distribution - USA (nationwide) and to the countries of : Canada, Australia, UAE, Kuwait, Saudi Arabia, South Africa
- Recall number
- Z-0026-2018
- Recall status
- Open, Classified
- CDRH record ID
- 158454
- FDA record created
- 2017-10-19
- Recall event number
- 77987