Skip to content

Public FDA record

Z-0026-2022: OPTETRAK Comprehensive Knee System including OPTETRAK Logic RBK PS Tibial Components (Product Line: 02-01

Recall record Z-0026-2022 for Exactech, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0026-2022
Company in FDA record
Exactech, Inc.
FDA classification date
Oct 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

Source facts

Action
Exactech notified distributors and sales representatives on about 08/30/2021 via letter titled "URGENT MEDICAL DEVICE RECALL." Actions included removing all Knee and Ankle UHMWPE products labeled with an 8-year shelf life and not packaged in EVOH/Nylon bags, in a phased approach over 12 months. Phase 1: immediately return all knee and ankle UHMWPE devices labeled with an 8-year shelf life that will be 5 years old or older by 08/31/2022 not packaged in EVOH/Nylon bags. Phase 2: between 05/31/2022 to 08/31/2022, returning all remaining knee and ankle UHMWPE devices labeled with an 8-year shelf life not packaged in EVOH/Nylon bags. A communication to healthcare professionals should follow. Instructions to recipients included to immediately identify and quarantine any affected product in inventory, notify customers down the line if further distributed, and complete and return the Recall Inventory Response Form to Exactech via email at recalls@exac.com. Exactech expanded their recall and sent an updated notification to US agents and distributors on 02/07/2022. On about 02/08/2022, a letter to surgeons and hospitals was sent. The notifications indicated that all non-conforming devices were included in the recall, to cease use of those devices, and work with sales agents to return any inventory. The letters to surgeons and hospitals also advised physicians regarding follow up with implanted patients.
Initiated
2021-08-30
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide distribution and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CYPRUS, ECUADOR, FRANCE, FRENCH POLYNESIA, GERMANY, GHANA, GREAT BRITAIN, GREECE, GUATEMALA, INDIA, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC OF, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, POLAND, PORTUGAL, ROMANIA, SINGAPORE, SPAIN, SWITZERLAND, SYRIAN ARAB REPUBLIC, THAILAND, TUNISIA, TURKEY, VENEZUELA .
Recall number
Z-0026-2022
Recall status
Open, Classified
CDRH record ID
189267
FDA classified
2021-10-04
Recall event number
88570

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗