Mislabeling. The ELAN 4 FIXED DURAGUARD STANDARD was labeled as "Long" and the ELAN 4 FIXED DURAGUARD LONG was labeled as "Standard."
Source facts
Action
Asculap notified consignees with an "Urgent Field Safety Notice" dated 8/18/2025 via email. Consignees were instructed to identify affected units, arrange for their return, and complete and return the response form provided with the letter.
If you have any question, call 484.240.8373 or email:allison.longenhagen@bbraunusa.com.
Initiated
2025-08-18
Root cause
Process control
Distribution
US distribution to Missouri. OUS International distribution to Norway, Portugal, France, Great Britian, Czech Republic, Germany, Turkey, Brazil, Japan, Chile, China, Hong Kong, Indonesia, India, Iran, Mexico, Malaysia, Philippines, Saudi Arabia, Thailand.
Recall number
Z-0026-2026
Recall status
Open, Classified
CDRH record ID
215819
FDA classified
2025-10-02
Recall event number
97507
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.