Open original FDA record ↗Recall record
- FDA identifier
- Z-0027-04
- Company in FDA record
- Instrumentation Laboratory Co.
- FDA classification date
- Oct 15, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Lactate Disposable Membrane Caps linearity was determined to be significantly biased and could potentially cause erroneous patient results above 6.0 mmol/L.
Source facts
- Action
- Instrumentation Laboratories notified two (2) U.S. direct customers on August 28, 2003 by telephone, regarding Lot N30601. Follow-up phoning emphasizing significance of linearity problem (and additional involved lot N307 - for 1 affected customer) performed on September 3, 2003. Phone call followed with faxback acknowledging notification and disposition of remaining product lot(s). IL service also dispatched to customer sites on September 4, 2003. International Distributor Notices sent to IL Europe and IL Mexico on September 4, 2003. Product Information Bulletins circulated on September 3, 2003, to IL personnel worldwide.
- Initiated
- 2003-08-28
- Root cause
- Other
- Terminated
- 2012-05-23
- Distribution
- IL, NY International: Italy, Mexico
- Recall number
- Z-0027-04
- Recall status
- Terminated
- CDRH record ID
- 29332
- FDA classified
- 2003-10-15
- Recall event number
- 27163