Z-0027-2024: Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016

Recall record Z-0027-2024 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0027-2024
Company in FDA record
Bard Peripheral Vascular Inc
FDA classification date
Oct 4, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly fitting into the coaxial cannula and accessing target tissue, which may lead to repeated procedure, and procedural complications, such as pain/discomfort, pneumothorax, and bleeding.

Source facts

Action
On 8/10/23, BD mailed or emailed recall notices to Medical Directors, Risk Managers, Medical Device Safety Offices, Lab Managers, Clinical Engineers, and Distributors who were told to do the following: 1) Discontinue distribution and use and immediately dispose of all affected product. 2) Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. 3) Distributors, identify all customers within your distribution network that purchased any affected product and provide a copy of the customer letter to all customers and advise them of this field action notification on behalf of the recalling firm. 2) Complete the attached Response Form and return via email to: BDRC45@bd.com If you require further assistance contact: the North American Regional Complaint Center: 1-844-8BD- LIFE (1-844-823-5433) M-F 8am - 5pm CT, or Email: productcomplaints@bd.com
Initiated
2023-08-10
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of PR, CO, TX, CA, WA, OR, AZ, NJ, NY, OH, NC, MD, MI, LA, IN, AK, TN and the countries of CA, AU, BE.
Recall number
Z-0027-2024
Recall status
Open, Classified
CDRH record ID
203206
FDA classified
2023-10-04
Recall event number
92994

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗