Skip to content

Public FDA record

Z-0027-2025: Aligned Medical Tonsil Tray REF AMS9300A that contains the Cardinal Health Salem Sump PVC tubes. Product

Recall record Z-0027-2025 for Windstone Medical Packaging, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0027-2025
Company in FDA record
Windstone Medical Packaging, Inc.
FDA classification date
Oct 8, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to convenience kits containing a kit component that is under a recall action for improper use.

Source facts

Action
On 08/16/2024, the firm email an "URGENT: MEDICAL DEVICE PRODUCT CORRECTION" Letter to customers informing them that Cardinal Health is conducting a product correction for its Salem Sump PVC Tubes. These Sump PVC tubes were included in the Aligned Medical convenience kits. The correction for this Recall is an updated product label, an electronic instruction for use (eIFU). The updated labeling is to address improper use of the Sump PVC tubes which can lead to breakage of the Salem Sump Anti Reflux Valve (ARV) that increases risk to patients. Customers are instructed to: 1. Review your inventory for the affected kits, all lots are impacted 2. Communicate the change to the use instructions with all personnel that utilize Salem Sump" products contained within AMS kits. 3. If you have further distributed this product, please identify your distribution partners, and notify them at once of the product correction. Your notification to your customers should include a copy of this notification letter. 4. Please complete and return the AMS Urgent Medical Device Product Correction Response Required form within 5 business days and return via email to fieldcorrectiveaction@alignedmedical.com even if you have no affected product on hand. For questions or concerns, contact Aligned Medical Solutions at fieldcorrectiveaction@alignedmedicalsolutions.com
Initiated
2024-08-16
Root cause
Use error
Distribution
U.S. Nationwide distribution in the states of CA, MO, OH, VA, and WY.
Recall number
Z-0027-2025
Recall status
Open, Classified
CDRH record ID
209952
FDA classified
2024-10-08
Recall event number
95306

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗