Patient samples with very high lactate values may falsely be reported as within the normal range without the instrument giving a warning flag to the technician.
Source facts
Action
Recall letters were sent to U.S. customers on 9/2/03 instructing them not to run lactate with these analyzers. An update notification letter was sent on or about 9/18/03 telling customers to discontinue use of the product for manual lactate determinations.
Initiated
2003-09-02
Root cause
Other
Terminated
2003-12-10
Distribution
United States, Canada and Germany
Recall number
Z-0028-04
Recall status
Terminated
CDRH record ID
29333
FDA classified
2003-10-15
Recall event number
27164