Potential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can
stretch into a configuration that blocks the flow of insufflation gas (C02).
Source facts
Action
Customers were notified via FedEx or UPS on June 25,2008 with an overnight letter. The letter advises that Applied Medical is conducting a voluntary recall of the 15mm Separator Access Systems, Models C0604, C0605, C0606 and C0607 due to the potential inability to insufflate through the stopcock. Per the letter, the firm is asking
that all 15mm product listed in the letter be returned. A response form is provided.
Initiated
2008-06-25
Root cause
Process control
Terminated
2012-03-26
Distribution
Nationwide and Canada distribution.
Recall number
Z-0028-2009
Recall status
Terminated
CDRH record ID
73514
FDA classified
2008-10-07
Recall event number
49445