Skip to content

Public FDA record

Z-0028-2022: TECNIS Eyhance IOLs are permanent intraocular implants intended for implantation in the capsular bag. Th

Recall record Z-0028-2022 for Johnson & Johnson Surgical Vision Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0028-2022
Company in FDA record
Johnson & Johnson Surgical Vision Inc
FDA record creation date
Oct 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Nonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal endotoxin specification limit and the FDA recommended endotoxin specification limit. Product was released in error based on the results of additional testing performed during the investigation into release testing nonconformance.

Source facts

Action
On 08/18/2021, the firm sent via FedEx standard overnight an "URGENT DEVICE Recall" Letter informing customers that it inadvertently release of nonconforming product (high endotoxin levels). Customers are instructed to: 1. Complete and return the Customer Reply Form to acknowledge receipt of the Recall Notice and email to RegCompliOne@its.jnj.com within 3 business days of receipt of the recall notice. 2. Review their patient records to confirm implantation of the impacted lens. 3. If no unexpected postoperative outcome occurred following implantation, no further action is required after Customer Reply Form has been returned. 4. If issues were encountered, inform the Recall Firm by calling 877-266-4543, Option 2. If reporting a complaint, provide the serial number, the date of surgery, a description of the event, and patient outcome. For any questions, contact Customer Support at; 877-266-4543, Option 2 or email RegCompliOne@its.jnj.com
Initiated
2021-08-06
Root cause
Process control
Terminated
2023-03-06
Distribution
U.S. Nationwide distribution in the state of IL.
Recall number
Z-0028-2022
Recall status
Terminated
CDRH record ID
189006
FDA record created
2021-10-04
Recall event number
88564

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗