Z-0029-2007: MRP-7000 and AIRIS Magnetic Resonance Imaging Systems, Software Versions: V7.0A to V7.0J.

Recall record Z-0029-2007 for Hitachi Medical Systems America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0029-2007
Company in FDA record
Hitachi Medical Systems America Inc
FDA classification date
Oct 12, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Image orientation error. When a 3D Maximum Intensity Projection (MIP) image data set is transferred from the MRI system to a computer workstation via the DICOM protocol, the anatomical markers do not change between the images as they rotate.

Source facts

Action
On 8/16/2006, Hitachi issued two (2) Device Correction letters. The first letter was addressed to the five (5) customers whose system units are currently ''de-installed''. This letter informs the customers of the problem and asks that the customer(s) please contact the firm''s Customer Service Center on their toll-free tel. # 1-800-800-4925, in order to schedule a device correction should the customer ever re-activate/install their system(s). The second letter, issued to the customers with activated MRP-7000 and/or AIRIS MRI systems, informs the customers of the problem, and states that a HMSA Service (person) will be contacting the customer in order to schedule a device correction which should address the problem.
Initiated
2006-08-16
Root cause
Other
Terminated
2012-04-13
Distribution
Nationwide and Puerto Rico.
Recall number
Z-0029-2007
Recall status
Terminated
CDRH record ID
48211
FDA classified
2006-10-12
Recall event number
36224

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗