Z-0029-2026: NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Recall record Z-0029-2026 for NOXBOX LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0029-2026
Company in FDA record
NOXBOX LTD
FDA classification date
Oct 6, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.

Source facts

Action
On September 9, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers: Actions to be taken: - Customers should ensure that all affected personnel are provided with a copy of the recommendations in this letter. - Users should avoid rapidly pressing buttons in quick succession (e.g., go to next screen, accept a value, change a value, backing out, repeating this) as it is necessary to allow an opportunity for the device to respond to user inputs before the next command to avoid a potential software interruption and subsequent restart. - NOxBOX Ltd. is asking customers in the U.S. to return the attached response form to Linde Gas & Equipment as instructed on the form. Recalling firm actions: NOxBOX Ltd. is in the process of developing a new software version that will include a modification to the software to resolve this issue, which will be released pending required regulatory review and approval. Once the software version is released, NOxBOX Ltd. will be back in contact with you to update your device software. If you have any questions about this notice or need assistance, please reach out to your local Clinical Educator or the Customer Excellence Center at 1-833-NOX-VENT (669-8368) or at LG.US.NOXIVENT@Linde.com.
Initiated
2025-09-09
Root cause
Software design
Distribution
Domestic: TN;
Recall number
Z-0029-2026
Recall status
Open, Classified
CDRH record ID
215836
FDA classified
2025-10-06
Recall event number
97516

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗