Skip to content

Public FDA record

Z-0030-2012: Minutex D-Dimer ; Trinity Biotech plc, Bray, Co. Wicklow, Ireland. Tel: (353) 1 276 9800, Fax: (353) 1 27

Recall record Z-0030-2012 for Tcoag Us, Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0030-2012
Company in FDA record
Tcoag Us, Incorporated
FDA classification date
Oct 7, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There is a performance issue with Minutex D-Dimer Kit which may cause false negative results for patient samples containing levels of greater than 250ng/mL D-Dimer.

Source facts

Action
The firm,Tcoag US, Incorporated, sent an "Urgent Device Recall" letter dated June 13, 2011 via UPS to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately examine their inventory and quarantine the products subject to recall; complete and return the attached Product Recall Notification Response Form by June 22, 2011 via fax at 973-775-3085; scrap the products consistent with their laboratory policy and if they have further distributed these products, notify their customers to quarantine and scrap the recalled product. For questions or technical assistance, please call the Tcoag hotline at 1-888-291-0415, option 2 (7.30 am to 6.00 pm Eastern Time).
Initiated
2011-06-13
Root cause
Nonconforming Material/Component
Terminated
2012-12-11
Distribution
Worldwide distribution: USA (nationwide) and country of: Canada.
Recall number
Z-0030-2012
Recall status
Terminated
CDRH record ID
104078
FDA classified
2011-10-07
Recall event number
59957

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗