Z-0030-2020: L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethyl

Recall record Z-0030-2020 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0030-2020
Company in FDA record
Baxter Healthcare Corporation
FDA record creation date
Oct 3, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may predispose the patient to peritonitis.

Source facts

Action
On 10/10/2018, and 10/15/2018, Urgent Product Recall notices were mailed to dialysis providers at medical centers. Customers were asked to do the following: 1. Locate and remove the affected product lot from your facility. The product code and lot number can be found on the individual product or shipping carton. 2. Contact recalling firm for Service to arrange for return and credit. Center for Service can be reached at 888-229-0001. 3. If you purchased this product directly from the recalling firm, complete and return the enclosed customer reply form. 4. If you purchased this product from a distributor, please contact your distributor for return and credit. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a consumer-level recall of the affected product that you distributed to customers. Customers with additional questions are encouraged to contact: Corporate Product Surveillance at 800-437-5176, Clinical Helpline at 888-736-2543, and Renal Therapy Services at 800-553-6898. On 10/15/2018, Urgent Product Recall notices were mailed to Peritoneal Dialysis Patients who were asked to do the following: 1. Locate and remove the affected product lot from your supplies. The product code and lot number can be found on the individual product or shipping carton. 2. Contact recalling firm to arrange for return and credit. HomeCare Services can be reached at 800-284-4060. 3. Complete and return the enclosed Home Patient Reply Form.
Initiated
2018-10-10
Root cause
Mixed-up of materials/components
Terminated
2020-07-22
Distribution
U.S.: VA, DC, CA, FL, SC, MD, DE, PA, NC, WA, GA, TX, NV. No foreign (OUS) distribution.
Recall number
Z-0030-2020
Recall status
Terminated
CDRH record ID
173968
FDA record created
2019-10-03
Recall event number
83191

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗