Open original FDA record ↗Recall record
- FDA identifier
- Z-0031-2009
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Oct 2, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Boston Scientific is initiating a voluntary recall of the Platinum Plus" Guidewire (peripheral). BSC has received reports that product may be labeled as a 260 CM long guidewire when the actual packaged device is a 180 CM long guidewire and the product may be labeled as a 180 CM long guidewire when the actual packaged device is a 260 CM long guidewire. There is no expected potential for injury to
Source facts
- Action
- A "Boston Scientific Urgent Voluntary Medical Device Recall or Correction - Immediate Action Required" letter dated 8/25/08 was sent to each affected account (To The Hospital Field Action Contact). The letter includes the customer steps for recall, a complete listing of all product batches/lots being recalled, the Reply Verification Tracking Form(s), and a return shipping label. Contact Boston Scientific Corporation at 1-763-494-1700 for assistance.
- Initiated
- 2008-08-25
- Root cause
- Labeling mix-ups
- Terminated
- 2011-12-10
- Distribution
- USA: CA FL GA IL IN LA MN MO NJ NV NY OH PA TN TX UT WA WI and WV OUS: France, Germany, Great Britain, Italy, and the Netherlands.
- Recall number
- Z-0031-2009
- Recall status
- Terminated
- CDRH record ID
- 73482
- FDA classified
- 2008-10-02
- Recall event number
- 49462