Z-0032-2014: FSE Image Upgrade Kit, Catalog # 92019C This kit is used to install and/or upgrade software used with th

Recall record Z-0032-2014 for Bio-Rad Laboratories. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0032-2014
Company in FDA record
Bio-Rad Laboratories
FDA classification date
Oct 21, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The FSE Image Upgrade kit (part 92019C) is recalled due to issue related to the export of data from the EVOLIS analyzer to Laboratory Information Systems (LIS) when using the MONOLISA Anti-HBs EIA Quantitative Determination in EVOLIS APF version 6.0.

Source facts

Action
BioRad has sent an Urgent Product Correction letter dated August 19, 2013, to customers who had the EVOLIS FSE Upgrade Image Kit installed on the EVOLIS at their site by a Bio-Rad Laboratories Field Service Engineer (FSE). This issue impacts laboratory only if you are running the MONOLISA Anti-HBs EIA Quantitative Determination method with EVOLIS APF version 6.0 and TimeLiner Data Management Software. If you use this Quantitative Determination method with EVOLIS APF version 6.0, please contact Bio-Rad Laboratories for additional information or assistance. Bio-Rad Laboratories is currently developing and qualifying a software update CD consisting of the EVOLIS APF V6.0b and Timeliner Data Management Software to correct the issue. The software update CD will be installed by Bio-Rad Field Service Engineers at each of the customer sites when installing the EVOLIS APF V6.0b. Customers should contact Bio-Rad Laboratories at 1-800-224-6723, Option 2 and then Option 3 for any questions about this product correction.
Initiated
2013-08-20
Root cause
Software change control
Terminated
2013-11-13
Distribution
US Distribution including the states of FL, IA, ME, MO, MS, NE, NY, TN, TX, and WA
Recall number
Z-0032-2014
Recall status
Terminated
CDRH record ID
122026
FDA classified
2013-10-21
Recall event number
66293

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗