Open original FDA record ↗Recall record
- FDA identifier
- Z-0032-2014
- Company in FDA record
- Bio-Rad Laboratories
- FDA classification date
- Oct 21, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
The FSE Image Upgrade kit (part 92019C) is recalled due to issue related to the export of data from the EVOLIS analyzer to Laboratory Information Systems (LIS) when using the MONOLISA Anti-HBs EIA Quantitative Determination in EVOLIS APF version 6.0.
Source facts
- Action
- BioRad has sent an Urgent Product Correction letter dated August 19, 2013, to customers who had the EVOLIS FSE Upgrade Image Kit installed on the EVOLIS at their site by a Bio-Rad Laboratories Field Service Engineer (FSE). This issue impacts laboratory only if you are running the MONOLISA Anti-HBs EIA Quantitative Determination method with EVOLIS APF version 6.0 and TimeLiner Data Management Software. If you use this Quantitative Determination method with EVOLIS APF version 6.0, please contact Bio-Rad Laboratories for additional information or assistance. Bio-Rad Laboratories is currently developing and qualifying a software update CD consisting of the EVOLIS APF V6.0b and Timeliner Data Management Software to correct the issue. The software update CD will be installed by Bio-Rad Field Service Engineers at each of the customer sites when installing the EVOLIS APF V6.0b. Customers should contact Bio-Rad Laboratories at 1-800-224-6723, Option 2 and then Option 3 for any questions about this product correction.
- Initiated
- 2013-08-20
- Root cause
- Software change control
- Terminated
- 2013-11-13
- Distribution
- US Distribution including the states of FL, IA, ME, MO, MS, NE, NY, TN, TX, and WA
- Recall number
- Z-0032-2014
- Recall status
- Terminated
- CDRH record ID
- 122026
- FDA classified
- 2013-10-21
- Recall event number
- 66293