- FDA identifier
- Z-0033-06
- Company in FDA record
- Baxter Healthcare Corp.
- FDA classification date
- Oct 8, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
If a pharmacy end user alters the Administration Unit of Measure (AUOM) of an ordered drug, the AUOM will display incorrectly on the Medication Administration Record for the patient, possibly leading to medication errors.
Source facts
- Action
- Urgent Device Correction recall letters dated 6/16/05 were sent to the direct accounts on the same date via first class mail. The letter describes the issue with the Administration Unit of Measure (AUOM), and informed the accounts that changing the AUOM of an ordered drug presents a patient safety issue resulting in a potential for a medication error. The accounts were requested to eliminate or minimize any further AUOM changes until a software upgrade is available. If an AUOM change is required the pharmacy end user must contact the Baxter Patient Care Helpdesk at 1-877-835-2727 prior to any changes being made. A Technical Bulletin was provided with the letter to guide the pharmacy end user through the process of an AUOM change with the aide of a Client Support Specialist.
- Initiated
- 2005-06-16
- Root cause
- Other
- Terminated
- 2008-02-06
- Distribution
- Missouri, Massachusetts, Vermont, California and Canada.
- Recall number
- Z-0033-06
- Recall status
- Terminated
- CDRH record ID
- 40109
- FDA classified
- 2005-10-08
- Recall event number
- 32449