Open original FDA record ↗Recall record
- FDA identifier
- Z-0033-2012
- Company in FDA record
- Isopure Corp
- FDA classification date
- Oct 11, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software problem causes transfer line between mixing and storage tank to not be rinsed after disinfection, leaving residual chlorine in the line.
Source facts
- Action
- Isopure sent a letter dated May 19, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Isopure devised an immediate solution which must be implemented at their faclity in conjunction with the disinfection procedure. Customers were asked to inform all appropriate personnel, clinical and technical, who may be involved in operating the system during the normal disinfection schedule. On 5/19/2011, the firm sent letters to their customers via US mail and email. They also telephoned their customers to notify them of the problem.
- Initiated
- 2011-05-19
- Root cause
- Software design
- Terminated
- 2013-02-19
- Distribution
- Nationwide Distribution including AR, CT, GA, MA, ME, NJ, PA, TN and WV.
- Recall number
- Z-0033-2012
- Recall status
- Terminated
- CDRH record ID
- 103921
- FDA classified
- 2011-10-11
- Recall event number
- 59879