Z-0033-2022: Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 07P5605) - Product Usage: use in the cal

Recall record Z-0033-2022 for Sentinel CH SpA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0033-2022
Company in FDA record
Sentinel CH SpA
FDA record creation date
Oct 6, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Calibrator vials were incorrectly labeled. Use of incorrectly labeled vials will cause the instrument to generate a warning that calibrators are incorrectly loaded and calibration will not be completed successfully.

Source facts

Action
Firm notified distributor July 29, 2021, and provided an Urgent Field Safety Notice letter to be forwarded to final end users on August 20, 2021. Customers were advised to do the following: - Immediately discontinue use of affected lot of calibrator - Destroy any remaining inventory of affected lot according to your laboratory procedures - Contact Abbott to order replacement material - Review the content of the communication with your Medical Director - Retain the letter for future reference Please forward notification to all individuals in your organization who need to be aware, or to any individuals or organizations where the recalled lot was transferred. If you have questions regarding this recall, please contact your local area Customer Service.
Initiated
2021-07-29
Root cause
Under Investigation by firm
Terminated
2023-07-24
Distribution
US Nationwide distribution in the states of LA, MD, MN, ND, NY, WV, and Puerto Rico.
Recall number
Z-0033-2022
Recall status
Terminated
CDRH record ID
189277
FDA record created
2021-10-06
Recall event number
88680

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗