Open original FDA record ↗Recall record
- FDA identifier
- Z-0033-2026
- Company in FDA record
- Medtronic MiniMed, Inc.
- FDA classification date
- Oct 2, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection.
Source facts
- Action
- On May 7, 2025, Medtronic issued a "Urgent Field Safety Notice" to affected consignees. Medtronic asked consignees to take the following actions: 1. Verify if the sensors you currently have are from the impacted lot. 2. If you have a Simplera" sensor from the impacted lot, do not use this Simplera" sensor and dispose of it according to local regulations. 3. Please contact our helpline/your Medtronic contact at 01923 205167 to request a replacement. 4. Please acknowledge that you have read and understood this notification and have followed the actions listed in this letter, even if your sensor is unaffected, or if you do not have any sensors from the affected lot.
- Initiated
- 2025-05-07
- Root cause
- Process design
- Distribution
- US: No distribution OUS: Austria, Belgium, Italy, Spain, Switzerland, United Kingdom
- Recall number
- Z-0033-2026
- Recall status
- Open, Classified
- CDRH record ID
- 213511
- FDA classified
- 2025-10-02
- Recall event number
- 96745