Philips Medical Systems North America Co. Phillips
FDA classification date
Oct 8, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Potential for Internal Sound Amplifier to malfunction which could delay treatment.
Source facts
Action
On 11/20/08, Philips began sending the URGENT-Medical Device Correction letter to their consignees via e-mails. On 11/21/08, the firm sent out the recall letter via United State Postal Service.
The letter states the affected device as Philips IntelliVue Information Centers, running on the HP rp5700 PC platform equipped with an active speaker solutions using a DB-15 to 3.5 mm cable set.
The consignees are advised to ensure that the audible alarm system is clear, free of noise and distortion and adjustable to different sound levels which is explained in p. 5-47 "Adjusting Alarm Tone Volume" of IntelliVue of the Instruction for Use, Release K Manual, supplied with the hardware, Philips Part Number 453564062301.
Consignees can call Philips Healthcare Customer Care Service Center at 1-800-722-9377, #3, #1 and reference FCO 99862012.
Initiated
2008-11-21
Root cause
Component change control
Terminated
2010-10-12
Distribution
Nationwide Distribution
Recall number
Z-0034-2011
Recall status
Terminated
CDRH record ID
74994
FDA classified
2010-10-08
Recall event number
50215