Skip to content

Public FDA record

Z-0034-2012: Axle Interspinous Fusion System Torque Driver Handle, Model #: X060-0323, Lot # 525941, S/Ns 001-017, Lot

Recall record Z-0034-2012 for X Spine Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0034-2012
Company in FDA record
X Spine Systems Inc
FDA classification date
Oct 11, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The firm learned that the Torque Driver Handles were not actuating at a proper torque. It was determined that the handles were speed determinant and could output a higher torque if the handles were turned at a very slow speed.

Source facts

Action
X-spine Systems, Inc. sent an URGENT DEVICE RECALL letter dated September 12, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their stock immediately to determine if they have any affected product. If so, customers were instructed once the set screw is tightened onto the locking plate, and resistance is detected in the handle, it will take less than 2 seconds to hear the handle "click." The "click" indicates the torque limit was reached. If further resistance is felt after 2 seconds, the surgeon should release the handle from tension, and try again. If the product was distributed further, they should contact their accounts, and advise them of this recall. The firm will send replacement handles. For any questions call (937) 847-8400, ext 115.
Initiated
2011-07-15
Root cause
Component design/selection
Terminated
2013-03-01
Distribution
Nationwide Distribution including CA, CO, FL, IA, IN, LA, NY, PA, TN, TX and UT.
Recall number
Z-0034-2012
Recall status
Terminated
CDRH record ID
103598
FDA classified
2011-10-11
Recall event number
59756

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗