Skip to content

Public FDA record

Z-0035-04: Neonatal/Infant Hyperinflation Circuit, consisting of a 0.5 L breathing bag, pressure monitoring tube, ox

Recall record Z-0035-04 for VIASYS Med Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0035-04
Company in FDA record
VIASYS Med Systems
FDA classification date
Oct 15, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The exhalation hole in the sideport relief valve may be blocked, allowing excess pressure build-up.

Source facts

Action
VIASYS telephoned the direct accounts and sent out follow-up faxes on 9/10-11/03, informing them of the potential for a blocked exhalation hole in the sideport relief valve cap which may result in excess pressure build-up. The accounts were instructed to check their inventory for the affected catalog numbers and lots numbers, place any found on hold for return to VIASYS, and sub-recall the affected product from their customers.
Initiated
2003-09-10
Root cause
Other
Terminated
2004-03-24
Distribution
Missouri, Texas, New York, Illinois, Kansas, California, Pennsylvania, Georgia, Arizona, Massachusetts, Maine, Oklahoma and Virginia.
Recall number
Z-0035-04
Recall status
Terminated
CDRH record ID
29502
FDA classified
2003-10-15
Recall event number
27242

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗