Skip to content

Public FDA record

Z-0035-2025: Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Cod

Recall record Z-0035-2025 for Philips North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0035-2025
Company in FDA record
Philips North America
FDA classification date
Oct 8, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.

Source facts

Action
An IMPORTANT PRODUCT NOTICE dated 4/19/23 was sent to customers. The actions that you as a customer can take to minimize the effect of the problem This communication should be shared with all clinical staff to review and understand. Place this Important Product Notice with the documentation of the PIC iX Instructions for Use. Additional Actions Dependent on Normal Workflow: 1. Workf/ow A: MX40 Routinely Cleared from Sector: If the MX40 is routinely cleared from the sector upon discharge of a patient then no additional action is required. 2. Workflow B: MX40 not Routinely Cleared from Sector: Option 1: Change Default Settings to Clear Sector Automatically The PIC iX can be setup to routinely clear the sector by executing the following: a. This can be set as the default as needed by selecting the "Clear Sector" selection and the "Clear unlocked Telemetry Devices from bed" in the "Configuration, Patient Management/Workflow", as highlighted in the image below: Option 2: Clear MX40 from Sector Manually The MX40 can be cleared from the sector manually by following the steps below: a. Identify the sector to be cleared from the PIC iX. These steps may be repeated for all sectors in use if they are not routinely cleared. b. Select the "Manage Patient" button and the "Manage Patient" window will display. c. Select the "Clear Sector'' button and the discharge window will display. d. Check the box next to the "Clear Sector'' option, as shown in the example Image 4 below: e. Selecting the "Cancel" button does not change the sector assignment. Finalize the process by selecting the " Discharge" button. 3. Workflow C: Clear Sector Option Not Available: In some configurations, the clear sector function will not be available. If this occurs, please contact the local response center for guidance on how to proceed. 4. The actions planned by Philips to correct the problem A Philips representative will reach out to you to arrange for a software upgrade to your device. This upgrade wi
Initiated
2023-04-19
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-0035-2025
Recall status
Open, Classified
CDRH record ID
209983
FDA classified
2024-10-08
Recall event number
95325

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗