Open original FDA record ↗Recall record
- FDA identifier
- Z-0036-2014
- Company in FDA record
- Xoran Technologies, Inc.
- FDA record creation date
- Jun 26, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A description of the defect in the product or the manner in which the product fails to comply with an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu
Source facts
- Action
- All of the MiniCAT CT Scanners subject to these complaints were serviced by Xoran technicians and the out-of-specifications devices were brought into specification. This was accomplished by either readjusting the settings or by installing a new component.
- Initiated
- 2010-07-27
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2014-09-11
- Distribution
- Worldwide Distribution - US nationwide
- Recall number
- Z-0036-2014
- Recall status
- Terminated
- CDRH record ID
- 122074
- FDA record created
- 2015-06-26
- Recall event number
- 66335