Z-0036-2024: Meera Mobile Operating Table-to support and position the patient immediately before, during and after sur

Recall record Z-0036-2024 for Getinge Usa Sales Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0036-2024
Company in FDA record
Getinge Usa Sales Inc
FDA classification date
Oct 4, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive) of max. 7 sec. may result in the following reasonably foreseeable injury or damage to the health of patients and/or users

Source facts

Action
Maquet GMBH / Getinge initiated Urgent Medical Device Correction letter on 9/6/23 via Fed'X. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any of the Meera Mobile Operating Table with the serial numbers listed in this notice. Maintain awareness on this notice and related actions until your operating table has been updated to ensure effectiveness of the corrective action. Please complete and sign the attached MEDICAL DEVICE CORRECTION- RESPONSE FORM (page 4) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to meera-unintendedmovement2023.sw.qrc@getinge.com or by faxing the form to 800-746-3592 Please forward this information to all current and potential Meera Mobile Operating Table users within your hospital / facility. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Getinge has developed a correction to address this issue. A Getinge Service representative will contact you to schedule the correction. This work will be done at no cost to your facility. If you have any questions, please contact your Maquet /Getinge representative Getinge Technical Support at (888) 943-8872 (press option 4, then option 2, then option 1) Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Initiated
2023-09-06
Root cause
Software Manufacturing/Software Deployment
Distribution
US Nationwide distribution.
Recall number
Z-0036-2024
Recall status
Open, Classified
CDRH record ID
203256
FDA classified
2023-10-04
Recall event number
93005

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗