Yoke may become displaced. Yoke index safety system could become compromised, allowing a user to attach the wrong gas type to the yoke connection.
Source facts
Action
Airgas issued a Medical Device Removal letter dated 8/6/09 to its customers informing them of the problem and the need to have a yoke back-up plate installed on the Manifold Block Assembly.
Initiated
2009-08-06
Root cause
Component design/selection
Terminated
2010-01-26
Distribution
Nationwide distribution.
Recall number
Z-0037-2010
Recall status
Terminated
CDRH record ID
84605
FDA classified
2009-10-14
Recall event number
53081