Skip to content

Public FDA record

Z-0037-2020: TxRx Knee 15 Coil 1.5T QED part number: Q7000057 Siemens Model # 10606829

Recall record Z-0037-2020 for Quality Electrodynamics LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0037-2020
Company in FDA record
Quality Electrodynamics LLC
FDA record creation date
Oct 3, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.

Source facts

Action
The firm sent a notification letter to their customers who that have received the affected device. Customers were instructed to. 1. Immediately examine your inventory and quarantine product subject to recall. 2. if you have further distributed this product to MRI end users, please identify your affected customers and notify them. 3. Please ensure they understand to stop use of the coil for scanning and to notify you immediately.
Initiated
2019-08-20
Root cause
Nonconforming Material/Component
Terminated
2020-05-26
Distribution
Worldwide distribution - US in the states: NC OUS: Germany
Recall number
Z-0037-2020
Recall status
Terminated
CDRH record ID
176331
FDA record created
2019-10-03
Recall event number
83775

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗