Tube may break. The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrectly. These tubes have the potential to break off at the black plastic connectors. The breakage may occur during the installation of the tubes or after a short time in use.
Source facts
Action
By telephone call or service representative visit beginning July 18, 2007, Consumers/End Users are asked to locate the analyzers and yearly service kits concerned and replace all pump tubes with new ones.
Initiated
2007-06-22
Root cause
Process design
Terminated
2012-04-25
Distribution
Product was distributed throughout the United States and Canada.
Recall number
Z-0038-2008
Recall status
Terminated
CDRH record ID
53878
FDA classified
2007-10-11
Recall event number
38511