Z-0038-2010: AGI-2238 Portable 4 Cylinder Manifold Cart, for connecting medical gas cylinders.

Recall record Z-0038-2010 for Airgas Puritan Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0038-2010
Company in FDA record
Airgas Puritan Medical
FDA classification date
Oct 14, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Yoke may become displaced. Yoke index safety system could become compromised, allowing a user to attach the wrong gas type to the yoke connection.

Source facts

Action
Airgas issued a Medical Device Removal letter dated 8/6/09 to its customers informing them of the problem and the need to have a yoke back-up plate installed on the Manifold Block Assembly.
Initiated
2009-08-06
Root cause
Component design/selection
Terminated
2010-01-26
Distribution
Nationwide distribution.
Recall number
Z-0038-2010
Recall status
Terminated
CDRH record ID
84606
FDA classified
2009-10-14
Recall event number
53081

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗