Open original FDA record ↗Recall record
- FDA identifier
- Z-0038-2015
- Company in FDA record
- DePuy Orthopaedics, Inc.
- FDA classification date
- Oct 8, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Specific lots are being recalled due to the potential for the tabs to break, leaving fragments in the patient.
Source facts
- Action
- On February 25, 2013, DePuy Orthopaedics, Inc. issued a device correction for the specific lots of the RECLAIM Distal Reamer Extension impacted by this recall. The letter stated the reason for the recall and swap out. A subsequent letter outlined a formal swap-out process of the affected reamer extensions as part of the device correction strategy. Newly designed devices are now available. Questions about device swap-out information may be directed to Kim Earle, Recall Coordinator, 574-371-4917; and Surgeon questions may be directed to DePuy's Scientific Information Office at 1-888-554-2482.
- Initiated
- 2014-09-03
- Root cause
- Device Design
- Terminated
- 2015-11-09
- Distribution
- Worldwide Distribution -- USA, including the states of ME, MD, PA, MA, FL, LA, IL, IA, OH, MN, TX, MT, WA, CA, HI, NV, AZ, KY, WI, CO, GA, MI, NC, VA, NY, TN, and IN; and, the countries of Canada, Australia, Austria, Belgium, Czech Republic, United Kingdom, France, Germany, Ireland, Israel, New Zealand, Norway, Poland, Slovenia, South Africa, Switzerland, and Chile.
- Recall number
- Z-0038-2015
- Recall status
- Terminated
- CDRH record ID
- 129656
- FDA classified
- 2014-10-08
- Recall event number
- 69104