Z-0038-2020: V-Twin, Catalog # 6002-800, UDI: 03661540600180 - Product Usage: An in vitro diagnostic device, which is

Recall record Z-0038-2020 for Vital Scientific N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0038-2020
Company in FDA record
Vital Scientific N.V.
FDA record creation date
Oct 4, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The action is being initiated because the cooling indicator board was improperly installed. This was first noticed by a service employee during assembly. If for some reason the cooling of the reagent rotor fails, the reagent cooling indicator light will not illuminate and the acoustic beeper will not go off. As a consequence the operator will not be warned and the reagent storage conditions may fall out of specification.

Source facts

Action
On 9/7/2018, a "Medical Device Recall" notice was sent to consignees. The recall notification, informed customers of the following: 1. ELITechGroup B.V. is requesting to correct this issue during the next preventive maintenance visit to your end user. The fix must be applied by a certified service engineer. No spare parts or special tooling is required to fix this issue. When done, this will be the long-term solution. See document 3084-249-00 Instructions Cooling indicator board for the detailed instructions. 2. Short term actions: ELITechGroup B.V. has already amended the applicable assembly and test instructions. No changes to the Service Manual are needed. Long term actions: Through this recall the applicable V-Twin analyzers in the field will be corrected. This action is expected to be finalized in September 2019. 3. For technical questions, please contact System Support department, tel. +31-313-430537 (Monday through Friday, 8:00-17:00 CET), e-mail service.ecsnl@elitechgroup.com On 9/13/2018, a revision to the "Medical Device Recall" notification was sent via email to consignees. The revised notification updated customers on the following issues. 1. Risk to Health: In the event there is no disturbance of the cooling of the reagent rotor the problem will not affect patients, health care providers, or other persons who are exposed to the device. The problem will not cause injuries or delays in treatment. In the event there is for some reason a disturbance of the cooling of the reagent rotor and this problem continued to remain undetected, the risk of an incorrect examination result and consequently an incorrect medical treatment and consequently harm to the patient is considered to be very low. Despite a possible failure of the reagent cooling unit, given the claimed on-board stability of the Siemens Healthcare Diagnostics Inc. Syva EMIT System Reagents for use on the V-Twin is typically four weeks, it is possible that all assays will co
Initiated
2018-09-07
Root cause
Device Design
Terminated
2022-06-29
Distribution
US Nationwide distribution including the state of New York.
Recall number
Z-0038-2020
Recall status
Terminated
CDRH record ID
176024
FDA record created
2019-10-04
Recall event number
83648

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗