Open original FDA record ↗Recall record
- FDA identifier
- Z-0039-2012
- Company in FDA record
- Richard-Allan Scientific Company
- FDA classification date
- Oct 12, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm has received two complaints of finger lacerations in operators of this device.
Source facts
- Action
- ThermoFisher Scientific sent "URGENT MEDICAL DEVICE CORRECTION, RE: Thermo Scientific HM355S-3 User Safety Advisory Notice" to all affected customers. The letter states the manufacturer identified a need to enhance the safety instructions and the description of the safety devices in the user manual. It mentions the HM3555S-3 User Manual has been modified to include more specific safe work practices and a new Quick -Reference Card summarizing the manual's safety enhancements and that the changes have been included on a CD that is enclosed with the letter. The firm requests confirmation of receipt of the new User Manual and Quick-Reference Card by October 31, 2011 to be sent by FAX to (269) 372-2674 For questions on this recall call the firms Technical Support representatives at 1-800-522-7270 08:00 AM- 4:30 PM.
- Initiated
- 2011-08-12
- Root cause
- Device Design
- Terminated
- 2013-07-25
- Distribution
- Worldwide Distribution: Nationwide distribution; including the country of Canada.
- Recall number
- Z-0039-2012
- Recall status
- Terminated
- CDRH record ID
- 102246
- FDA classified
- 2011-10-12
- Recall event number
- 59316