Skip to content

Public FDA record

Z-0039-2024: Citrine QTRAP Mass Spectrometer, Part Number 5063685

Recall record Z-0039-2024 for AB Sciex, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0039-2024
Company in FDA record
AB Sciex, LLC
FDA classification date
Oct 5, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Source facts

Action
An URGENT: RUO and MEDICAL DEVICE FIELD ACTION notification letter dated 7/20/23 was sent to customers. Actions to be taken by the Customer/User: SCIEX will provide a replacement nut for eligible sources and systems. Instructions will be provided to guide you through the replacement of the nut. The replacement parts with installation instructions are estimated to begin shipping in October 2023. SCIEX is currently notifying affected customers of the upcoming countermeasure. Upon availability, SCIEX will send you the replacement nut with instructions on how to install it on the ion source probe. We sincerely apologize for any inconvenience that this may cause you. SCIEX aims to provide you with products of the highest quality. If you have any questions regarding this notice, please contact https://sciex.com/aboutus/ contact-us or https://sciex.com/support/request-support.
Initiated
2023-08-30
Root cause
Component design/selection
Distribution
US Nationwide.
Recall number
Z-0039-2024
Recall status
Open, Classified
CDRH record ID
202634
FDA classified
2023-10-05
Recall event number
92810

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗