Z-0039-2025: GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model/Catalog/Code Numbers: a) 2089507

Recall record Z-0039-2025 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0039-2025
Company in FDA record
GE Healthcare
FDA classification date
Oct 9, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) version v6.0 SP11.x where the latest report is not shown by default to the user. The issue may occur when a study has multiple DICOM structured reports (i.e. preliminary, final, and/or addendum) stored in different series.

Source facts

Action
GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/29/2024 via mail. The notice explained the issue with the device, safety issue, and provided the following: "Actions to be taken by Customer/User: You can continue to use your device by following the recommendation below. To find the latest report, review all the reports available for the study in the Series Selector. Navigation to the series selection can be found in the User Manual under Section 2.4.3 of The Series Selector, with the subheading of Open a report/document. Please ensure all potential users in your facility are made aware of this safety notification and the recommended action." GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2024-08-29
Root cause
Process control
Distribution
Worldwide distribution.
Recall number
Z-0039-2025
Recall status
Open, Classified
CDRH record ID
209957
FDA classified
2024-10-09
Recall event number
95307

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗