Unanticipated technique for a cervical screw insertion caused the tip of the threaded inner shaft to break off.
Source facts
Action
Genesys Spine initiated their field action on November 15, 2015, by letter. The letter explained how and why the instrument failed and provided updated instructions on how the instrument was to be used. Consignees were provided the option to exchange the unit for another model. The firm expanded the action on September 22, 2017, to request the return of the product. On September 27, 2017, the firm sent a notice to three distributors who were not identified in the original group of consignees.
For further questions, please call (512) 381-7070.
Initiated
2015-11-05
Root cause
Component design/selection
Terminated
2018-07-19
Distribution
US Distribution to the states of : FL, TX, NY, IL, GA, AL, MA, MO
Recall number
Z-0040-2018
Recall status
Terminated
CDRH record ID
158848
FDA record created
2017-10-19
Recall event number
78181