Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate.
Source facts
Action
The service report in response to this complaint stated that the Head select relay board, chip and fan were replaced. The machine (ENTOOOO I) was subsequently tested. All tests were successful.
Initiated
2009-12-14
Root cause
Radiation Control for Health and Safety Act
Terminated
2014-09-11
Distribution
Worldwide Distribution - US nationwide
Recall number
Z-0041-2014
Recall status
Terminated
CDRH record ID
122083
FDA record created
2015-06-26