Open original FDA record ↗Recall record
- FDA identifier
- Z-0041-2017
- Company in FDA record
- Trumpf Medical Systems, Inc.
- FDA classification date
- Oct 5, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.
Source facts
- Action
- Trumpf Medical sent an Urgent Field Safety Notice dated July 8, 2015, to all affected consignees. The letter indicated that the tables should not be used until an authorized service engineer has exchanged the spindles of the affected operating tables. The letter included an acknowledgement form which is to be returned. The letter also requested a sub-recall if the product had been further distributed. For questions regarding this recall call 843-329-0543.
- Initiated
- 2016-07-10
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-03-02
- Distribution
- Worldwide Distribution - US including CA, CT, GA, IL, NC, OR, SC, VA, WI, WY, and Internationally to Australia, Azerbaijan, Brazil, Finland, France, Germany, Hungary, Italy, Luxembourg, Norway, Poland, Russia, Slovakia, and Switzerland
- Recall number
- Z-0041-2017
- Recall status
- Terminated
- CDRH record ID
- 149390
- FDA classified
- 2016-10-05
- Recall event number
- 75115