Z-0041-2020: Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer

Recall record Z-0041-2020 for Ohmeda Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0041-2020
Company in FDA record
Ohmeda Medical
FDA record creation date
Oct 4, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.

Source facts

Action
On October 5, 2018, the recalling firm notified affected customers of the product issue via an Urgent Medical Device Correction letter. Customers were instructed to do the following: You can continue to use your Giraffe OmniBed systems in open Bed warmer mode and incubator mode. Care should be taken to achieve visible cleanliness during the cleaning process. Design of the Giraffe OmniBed allows for your facilities normal cleaning to be performed, following the step by step procedure to remove the canopy seal to reach visible debris accumulation. Contact your Biomedical department to perform the removal and reassembly of the original seal fasteners, per the attached addendum, prior to performing the cleaning procedure. Attention to cleaning technique by personnel involved in the cleaning process will reduce the likelihood of occurrence of the mentioned condition. "Carefully follow the additional instructions provided to ensure continued safe use of the device and place the addendum in your User Manual. "Always follow your facility guidelines for cleaning and infection control. GE Healthcare will provide a Giraffe OmniBed Cleaning & Care Guidelines poster for display by your system. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2018-10-05
Root cause
Labeling design
Terminated
2020-06-02
Distribution
Worldwide distribution- Domestic nationwide and foreign distribution.
Recall number
Z-0041-2020
Recall status
Terminated
CDRH record ID
168671
FDA record created
2019-10-04
Recall event number
81419

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗