Open original FDA record ↗Recall record
- FDA identifier
- Z-0041-2020
- Company in FDA record
- Ohmeda Medical
- FDA record creation date
- Oct 4, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.
Source facts
- Action
- On October 5, 2018, the recalling firm notified affected customers of the product issue via an Urgent Medical Device Correction letter. Customers were instructed to do the following: You can continue to use your Giraffe OmniBed systems in open Bed warmer mode and incubator mode. Care should be taken to achieve visible cleanliness during the cleaning process. Design of the Giraffe OmniBed allows for your facilities normal cleaning to be performed, following the step by step procedure to remove the canopy seal to reach visible debris accumulation. Contact your Biomedical department to perform the removal and reassembly of the original seal fasteners, per the attached addendum, prior to performing the cleaning procedure. Attention to cleaning technique by personnel involved in the cleaning process will reduce the likelihood of occurrence of the mentioned condition. "Carefully follow the additional instructions provided to ensure continued safe use of the device and place the addendum in your User Manual. "Always follow your facility guidelines for cleaning and infection control. GE Healthcare will provide a Giraffe OmniBed Cleaning & Care Guidelines poster for display by your system. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Initiated
- 2018-10-05
- Root cause
- Labeling design
- Terminated
- 2020-06-02
- Distribution
- Worldwide distribution- Domestic nationwide and foreign distribution.
- Recall number
- Z-0041-2020
- Recall status
- Terminated
- CDRH record ID
- 168671
- FDA record created
- 2019-10-04
- Recall event number
- 81419