Z-0041-2024: Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product

Recall record Z-0041-2024 for Blue Belt Technologies, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0041-2024
Company in FDA record
Blue Belt Technologies, Inc
FDA classification date
Oct 5, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis

Source facts

Action
Smith & Nephew, Inc. (Blue Belt Technologies) issued Urgent Medical Device Correction Notice (C-2023-08) on September 7, 2023 via overnight mail. Letter states reason for recall, health risk and action to take: Required Actions for user/hospital: 1. Complete Correction Notice Acknowledgement Form and submit to FieldActions@smith-nephew.com. 2. Prior to use of software, contact your sales representative to update settings. 3. Upon completion of settings update, you may use the CORI Surgical System until the software update is deployed. 4. A Smith & Nephew Robotics Representative will be deployed to your site to implement a new software version to correct the issue once it is available. Email: FieldActions@smith-nephew.com Actions for Smith & Nephew Representatives 1. Work with customer to update CORI settings to prevent incorrect configuration. 2. Complete the attached Correction Notice Acknowledgement Form. 3. Upon completion of software upgrade, submit CORI In-Field Service Record to Field Action Team.
Initiated
2023-09-07
Root cause
Software Design Change
Distribution
US Nationwide distribution.
Recall number
Z-0041-2024
Recall status
Open, Classified
CDRH record ID
203324
FDA classified
2023-10-05
Recall event number
93046

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.