Skip to content

Public FDA record

Z-0041-2025: Inspire Model 4063 Stimulation Lead, component of the Inspire Upper Airway Stimulation (UAS) system.

Recall record Z-0041-2025 for Inspire Medical Systems Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0041-2025
Company in FDA record
Inspire Medical Systems Inc.
FDA classification date
Oct 9, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Labeling error with a mismatch between the shelf box serial number label and the serial number of the lead provided in the box.

Source facts

Action
Inspire notified consignees on 07/31/2024, via emailed letter. Consignees were instructed to confirm the serial number on the lead is the serial number used to register the patient and to contact your Inspire representative if any labeling inconsistency is identified for complaint reporting.
Initiated
2024-07-31
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Belgium, Switzerland, Germany, United Kingdom, Netherlands, Singapore.
Recall number
Z-0041-2025
Recall status
Open, Classified
CDRH record ID
209780
FDA classified
2024-10-09
Recall event number
95230

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗