Open original FDA record ↗Recall record
- FDA identifier
- Z-0041-2026
- Company in FDA record
- Fresenius Medical Care Holdings, Inc.
- FDA classification date
- Oct 3, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
To provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the dialyzer to securely affix the cap.
Source facts
- Action
- On August 29, 2025 IMPORTANT MEDICAL DEVICE INFORMATION letters were sent to customers. Actions to be taken: 1. Follow the New Instructions for Use provided in this letter when attaching the dialyzer caps. 2. Please report any complaints or adverse events to product.complaints@fmcna. com or FDA MedWatch at https://www.fda.gov/safety/medical-product-safetyinformation/ medwatch-forms-fda-safety-reporting. 3. Please forward this notice to any providers who may have received these products from you. Place this notification with the device to ensure awareness. For product technical questions or concerns, please contact the Technical Support at 1- 800-323-2732 or Customer Support at 1-800-323-5188. For any clinical or medical questions, please contact the Global Medical Information and Education Office at 1-855- 616-2309. This notification serves as important medical device information. FMCRTG is not requesting the return of any product. FMCRTG is working diligently to provide the updated Instructions for Use with the product and expects release in Q4 2025. The health and safety of your patients continue to be our highest priority. Thank you for your cooperation.
- Initiated
- 2025-08-29
- Root cause
- Device Design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0041-2026
- Recall status
- Open, Classified
- CDRH record ID
- 215902
- FDA classified
- 2025-10-03
- Recall event number
- 97543