Skip to content

Public FDA record

Z-0041-2026: Optiflux High Flux E-beam Dialyzer Model Numbers: (1) 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, (2) 05

Recall record Z-0041-2026 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0041-2026
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Oct 3, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

To provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the dialyzer to securely affix the cap.

Source facts

Action
On August 29, 2025 IMPORTANT MEDICAL DEVICE INFORMATION letters were sent to customers. Actions to be taken: 1. Follow the New Instructions for Use provided in this letter when attaching the dialyzer caps. 2. Please report any complaints or adverse events to product.complaints@fmcna. com or FDA MedWatch at https://www.fda.gov/safety/medical-product-safetyinformation/ medwatch-forms-fda-safety-reporting. 3. Please forward this notice to any providers who may have received these products from you. Place this notification with the device to ensure awareness. For product technical questions or concerns, please contact the Technical Support at 1- 800-323-2732 or Customer Support at 1-800-323-5188. For any clinical or medical questions, please contact the Global Medical Information and Education Office at 1-855- 616-2309. This notification serves as important medical device information. FMCRTG is not requesting the return of any product. FMCRTG is working diligently to provide the updated Instructions for Use with the product and expects release in Q4 2025. The health and safety of your patients continue to be our highest priority. Thank you for your cooperation.
Initiated
2025-08-29
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-0041-2026
Recall status
Open, Classified
CDRH record ID
215902
FDA classified
2025-10-03
Recall event number
97543

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗