Open original FDA record ↗Recall record
- FDA identifier
- Z-0043-2018
- Company in FDA record
- MicroPort Orthopedics Inc.
- FDA record creation date
- Oct 11, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The item is being recalled due to an inadequate pouch seal. This defect has the potential to lead to a loss of sterile pouch integrity, which could impact the products sterility assurance and may lead to serious adverse events such as infections, both localized and systemic.
Source facts
- Action
- Microport initiated their recall by letter sent via email on 09/18/2017. The letter identified the affected product, problem and actions to be taken. The consignees were directed to check their stock, including all kits and consignment stock to immediately determine if any affected product was in stock. The firm requested return of the recalled units. Consignees with questions regarding the return of the product were directed MPOs Customer Experience Department available Monday through Friday 7:30 am to 7:30 pm CST, at 1-866-872-0211.
- Initiated
- 2017-09-18
- Root cause
- Packaging process control
- Terminated
- 2021-04-08
- Distribution
- Worldwide Distribution - US Nationwide in the states of TX, AL, AZ CO, KS ,ME MO, WB, LA, CA, MD, including PR and the country of CANADA
- Recall number
- Z-0043-2018
- Recall status
- Terminated
- CDRH record ID
- 158776
- FDA record created
- 2018-10-11
- Recall event number
- 78150