Z-0043-2025: NxTAG Respiratory Pathogen Panel +SARS-CoV-2, REF: I056C0468

Recall record Z-0043-2025 for Luminex Molecular Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0043-2025
Company in FDA record
Luminex Molecular Diagnostics Inc
FDA classification date
Oct 9, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.

Source facts

Action
On 9/5/24, recall notices were sent to customers and distributors were were asked to do the following: 1. Two options are available to overcome the sporadic potential risk of the hMPV false positive result: A. Continue to use the product as per its instruction for use: in case of a positive result for hMPV, consider performing confirmatory testing with another method before reporting the hMPV result, if medically indicated. B. Exclude the hMPV from the reporting by removing this target from the assay analysis, as explained in the package insert under the chapter "Define Test Panels in SYNCT Software". See Document No. and Revision/Page reference: MLD-051-KPI-001 Rev H/Page 17, MLD-056-KPI-001 Rev. E/Page 18. 2. Notify others within your network that may have remaining inventory of those kits. 3. If you are a distributor, provide a copy of this letter to customers or organizations who received impacted lots. 4. Complete and return the response form via email to support@luminexcorp.com 5. For complaints, questions, or concerns, please contact Technical Support using the contact information: 1-877-785-2323 (U.S. and Canada). +1-512-381-4397 (Outside U.S. and Canada). 1-512-219-5114 (Fax). support@luminexcorp.com www.diasorin.com
Initiated
2024-09-05
Root cause
Device Design
Distribution
U Nationwide distribution in the states of OH, CA, NC, WI, TN, TX, MI, NY, MO, FL, MA, NJ, PA, VA, AZ, IL, OK, CO, MN.
Recall number
Z-0043-2025
Recall status
Open, Classified
CDRH record ID
210097
FDA classified
2024-10-09
Recall event number
95377

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗