Open original FDA record ↗Recall record
- FDA identifier
- Z-0044-2020
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA record creation date
- Oct 4, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
AssayTips part of the AssayTip/AssayCup Tray Tips used on the cobas e 801 module- may have an Abnormal Internal Structure and lead to Incorrect patient results
Source facts
- Action
- Roche Diagnostics issued Urgent Medical Notification letter to Distributor (TP-00744) and end users( TP-00741) on 8/13/19. Letter identifies problem, health risk and Actions required per the UMDC: " Check to see if you have any of the affected lot number(s) of AssayTips/AssayCup Trays on your cobas e 801 module or in your laboratory stock. (See pictures for lot number locations on packaging.). " Stop using the affected lot number(s) of AssayTips/AssayCup Trays immediately. " Discard the affected lot number(s) of AssayTips/AssayCup Trays according to your local waste management guidelines. To receive replacement product, fill out the attached fax form and fax or email it according to the instructions on the form. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions
- Initiated
- 2019-08-13
- Root cause
- Process change control
- Terminated
- 2022-06-30
- Distribution
- Nationwide
- Recall number
- Z-0044-2020
- Recall status
- Terminated
- CDRH record ID
- 176020
- FDA record created
- 2019-10-04
- Recall event number
- 83646