Open original FDA record ↗Recall record
- FDA identifier
- Z-0045-2007
- Company in FDA record
- Eatonform Inc
- FDA classification date
- Oct 19, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Certain lots of packet components of the firm's Doc-U-Dose Prescription Management System are separating from the paper along the sealed seam at the bottom of the packets causing the medications inside the separated compartments to fall completely out of the packet, or to become mixed into the wrong compartment of the packaging.
Source facts
- Action
- The firm issued recall notifications which were sent via UPS overnight delivery to all of their pharmacy customers on 8/2/2006, 8/4/2006, and 8/11/2006. In their recall notifications, the firm requested that the pharmacy customers: 1) cease using and distributing all Doc-U-Dose packets identified with the recalled lot numbers; 2) segregate and secure all Doc-U-Dose packets identified with the suspect lot numbers on-hand; 3) inform the firm as to whether or not they have any of the suspect product which they would be returning and the quantity of product to be returned; 4) return all packets to Eatonform at the recalling firms expense; and 5) notify their subaccount customers and ask that they cease using the device until such time as the problem is corrected.
- Initiated
- 2006-08-02
- Root cause
- Other
- Terminated
- 2012-05-21
- Distribution
- Nationwide
- Recall number
- Z-0045-2007
- Recall status
- Terminated
- CDRH record ID
- 47976
- FDA classified
- 2006-10-19
- Recall event number
- 36103